53
AUT
OCLA
VE SERIE SK07
TEST ProGrAMS SELECTIoN
By selecting the “test” sub-menu you access the list of test cycles that autoclave must execute:
1 - Vacuum Test
2 - 134 Helix Test
3 - 121 Helix Test
4 - Bowie&Dick Test
5 - Exit
The selection indicator is positioned on “Vacuum Test”. Select the desired program by moving
with the (+) and (-) keys and confirm with .
VACUUM TEST (Leakage test)
To execute this function the chamber must be cold and contain no instruments.
BoWIE & DICK TEST
There must be no tools in the chamber to carry out this test. The Bowie & Dick “package” to
test the porous load must be positioned on the lower tray.
HELIX TEST
There must be no tools in the chamber to carry out this test. The control tool to test the “Hollow
A” load must be positioned on the lower tray.
oPTIoNS SELECTIoN:
By selecting the “Options” sub-menu you access the list of configuration sub-menus for the
machine:
1 - Configuration
2 - Fill water
3 - Print cycle
4 - Delete memory
5 - Exit
Configuration sub-menu:
By selecting the “Configuration” sub-menu you access the list of configuration options for
the machine:
1 - Basic
2 - Advanced
3 - Special
4 - Exit
Basic selection:
By selecting the “Basic” sub-menu you access the list of configurations for the machine:
1 - Language
2 - Date
3 - Time
4 - Exit
Advanced selection:
Only an authorised technician can access this menu. By selecting the “Advanced” sub-menu
you access the list of configurations for the machine:
1 - Fill water options
2 - Saving
3 - Exit
Special selection:
Only an authorised technician can access this menu. By selecting the “Special” sub-menu you
access the list of configurations for the machine:
1 - Environmental Pressure
2 - LCD Contrast
3 - Password Management
4 - Exit