61
61
AUT
OCLA
VE SERIE SK07
MACHINE F
AUL
TS
Per
r
esettar
e il guasto e ripristinar
e il sistema eseguir
e la seguente pr
ocedura:
Premere contemporaneamente i tasti “
Enter
”, “+” e “-” ed attender
e il termin
e della pr
ocedura.
COD.
MESSAGE
DESCRIPTION
ACTION
AL001
Chamber
Pr
obe Fault
chamber probe open
Contact Faro technical support.
AL002
Band Pr
obe Fault
chamber probe in short circuit
Contact Faro technical support.
AL003
Guasto Sonda Fascia
band probe open
Contact Faro technical support.
AL004
Guasto Sonda Fascia
band probe in short circuit
Contact Faro technical support.
AL005
W
ater
Fill. T
imeout
maximum time exceeded to fill water
Check filling tube and presence of water in can.
AL006
Pr
essur
e T
ransducer
pressure transducer fault
Contact Faro technical support.
AL007
Door
Micr
o
unplanned door micro contact opening
Contact Faro technical support.
AL008
Safety Micr
o
unplanned door safety micro contact opening
Contact Faro technical support.
AL009
Blackout
blackout
Check if shut-of
f due to electrical system.
AL010
Safety Pr
essur
e
pressure over safe limit
Contact Faro technical support.
AL01
1
Corrupt Memory
corrupt memory
Contact Faro technical support.
AL012
Safety Micr
o
No door safety inserted
Contact Faro technical support.
AL013
PUSH Err
or
Software error
Contact Faro technical support.
AL014
POP
Err
or
Software error
Contact Faro technical support.
AL015
Reset Pr
essed
Reset key pressed
Proceed with autoclave release procedure.
AL016
Safety Micr
o
No door safety disconnection
Contact Faro technical support.
GUIDE To ProBLEMS