4000 Series
Page
34
of
41
3460006-17_serial series 5000000
Remote Running Contacts
Figure 9.2 – Remote contact terminal box with cover removed
Door open alarm – horn and strobe (optional)
In the event of a door opening while a profile or defrost cycle is running, there is a horn and strobe
light installed on the top of the unit that will activate and remote contact terminals 5 & 6 will close. See
the chart below for further explanation.
4000 series Unit State - Door Open
Alarm Contacts
Door
Position
Strobe/Horn
State
Contacts
5,6 NO
Unit turned on - In standby "off" state
CLOSED
OFF
OPEN
OPEN
OFF
OPEN
Unit turned on - Profile 1, Profile 2, or Profile 3 running
CLOSED
OFF
OPEN
OPEN
ON
CLOSED
Unit turned on - Defrost cycle running
CLOSED
OFF
OPEN
OPEN
OFF
OPEN