4000 Series
Page
38
of
41
3460006-17_serial series 5000000
11.0 General Maintenance
Model 4000 series GENERAL MAINTENANCE
Unit must be turned OFF & Unplugged during any and all maintenance/ service
.
Hazards!
Periodic Cleaning
Beginning with the initial installation, the interior surfaces of the cabinet should be periodically wiped
down with a solution of warm water and laboratory detergent. This solution will remove any odors
from spillage that has occurred. The detergent should be removed using distilled water followed by
a wipe down with isopropyl alcohol. The exterior of the cabinet should also be cleaned frequently
with a laboratory detergent solution followed with a distilled water rinse.
All surfaces (Internal and/or external) can be cleaned using isopropyl alcohol.
NOTE:
All moving parts have been permanently lubricated and will generally require no maintenance.
Cleaning the condenser (Air cooled only 4002/4005)
For air cooled units, monthly cleaning of the condenser will aid the heat transfer characteristics of the
refrigeration system and increase its efficiency. Dust, dirt, and lint will tend to accumulate on the fins of the
condensing unit. This obstruction will affect the flow of air through the condenser, thereby, lowering the
efficiency of the system.
a.
Remove condenser cover grill (Total 16 screws)
b.
Using a wire brush or a brush with stiff bristles can be used to loosen these particles that are
attached to the fins so that they may be removed with a vacuum cleaner. Failure to keep the condenser coil
clean and free of obstructions could result in temperature loss and damage to the compressor.
3. Re-install condenser cover grill (Total 16 screws)
Gasket maintenance
Periodically check the gaskets around the door for punctures or tears. Leaks are indicated by a streak of
frost which forms at the point of gasket failure. Make sure that the cabinet is level. Keep the door gaskets
clean and frost free by wiping gently with a soft cloth.