Model ULC Series
06012021ULC,Rev i
Page 15 of 54
b. Set-up
Screen
RETURN TO HOME SCREEN
RETURN TO PREVIOUS SCREEN
GO TO NEXT SCREEN IN THIS SEQUENCE
ALARM HIGH
: Touch the “Alarm High” display. A numerical entry keypad will be displayed. Enter
the high temperature alarm set point. The available range is +30°C to -80°C. The Factory Default
value is 30°C. Press ENT (enter) to save.
TEMP SET POINT
: Touch the “Temp Set Point” display to enter the chamber control temperature.
The available ranges are
ULC, ULCi -40°C to -80 °C
ULC-WR, ULCi-WR -20°C to -80°C
The Factory Default setting is -80°C.
Press ENT (enter) to save.
ALARM LOW
: Touch the “Alarm Low” display to enter the low temperature alarm set point. The
available range is -40°C to -99°C. The Factory Default is -88°C. Press ENT (enter) to save.
Home Screen
Each screen will display
the current internal
chamber temperature
Previous screen
Next Screen