Model ULC Series
06012021ULC,Rev i
Page 18 of 54
Press
to go to the next screen.
SETUP 4:
SINGLE/DUAL SYSTEM:
Touch either “Select Single Operation” or “Select Dual Operation”,
denoting if there are one or two refrigeration systems operational on this chamber. The software
to automatically switch between the two systems.
OTHER SIDE DETECTED:
This icon indicates if the controls are detecting one or two
refrigeration systems operational.
The opposite side system is not detected.
The opposite side system is detected.
SCREEN BRIGHTNESS:
Touch the “Screen Brightness” Arrow and the adjustment screen will
open. Adjust the display brightness to the desired level. Touch the ‘x’ in the upper right corner of
the screen to return to the System Configuration screen.
IGNORE DOOR SWITCH:
Normally, opening either door will cause the refrigeration system to
turn off and remain off as long as the door is open. This icon allows the operator to disable the
door switch input, so that the refrigeration system remains on when the door is open.
Note: this is not the recommended mode of operation.
The Ignore Door Switch mode should
only be used if a door switch fails. In this mode, the refrigeration systems will continue to operate
when the door is open or closed. Once the door switch is repaired, the Ignore Door Switch should
be turned off.
Pressing the icon will toggle between the two modes of door switch control.
The door switches will be monitored. Both refrigeration systems will turn off as
long as the door remains open.