7.0 Sensor Validation
Each system has an independent Control/display sensor (100 OHM 3 wire Platinum RTD, +/- 0.15
°C). Calibration may be performed using the Calibration screens on Technical Screen 3.
8.0 Alarms
Remote alarm contacts terminal block
The ULC is supplied with remote alarm signaling dry contacts (non-powered). Connections are
made to a terminal strip located on the right side of the unit near the top
.
Each pair of contacts
are electrically isolated from the other contacts. There are two sets of dry contacts. One set
activates when either door is open, the other activates for any alarm condition.
Both sets of contacts are normally open when the chamber is off.
Remote Alarm contacts in enclosure