Ripple
: this procedure provides the difference between maximum and minimum emitted light readings. This
value must be less than the hysteresis value
9.1.3
REFRIGERATION UNIT
Internal temperature
: sets the lower internal cabinet temperature of the machine, i.e. the temperature below
which the compressor is turned off. The pre-set value is +3°C. This value can be varied b2 to
+25°C.
D
efrost termination temperature:
every time the compressor pauses, the temperature of the evaporator is
controlled by the evaporator feeler. This command allows the evaporator temperature to be set, above which,
after dripping, the compressor can restart operating; this is to ensure complete defrosting of the evaporator
itself. The command is pre-set at 3°C and can vary between 3°C and 12°C.
Internal temperature
Defrost termination temperature
Defrost interval
Refrigerator safety action delay
Refrigerator safety temperature
Exit
Drawer safety 1
Drawer safety 2
Drawer safety 4
Drawer safety 5
Drawer safety 6
Drawer safety 7
Drawer safety 3