INSTALLATION
MT86/MT86HT Flow Meter
6
Fluid Components International LLC
Table 1 – Interconnecting Cable Minimum Conductor Size (Maximum AWG Number)
Connection
Maximum Distance for AWG
10 ft. (3m) 50 ft. (15m) 100 ft. (31m) 250 ft. (76m) 500 ft.
152m)
1000 ft.
(305m)
AC Power
22
22
22
20
18
16
Relay (2A)
28
22
20
16
12
10
Relay (10A)
22
16
12
8
6
Not
Recommend
Flow Element Wires*
24
24
24
22
22
18
*Note: Requires a shielded cable. The shield is connected to the GND pin of TS_ of the input board. The
other end of the shield is left floating (no connection to the flow element enclosure).
The standard cable provided is 22 AWG. Refer to
, page 35, for specific information.
Flow Transmitter Wiring Connections
, page 7, for the flow element wiring to the input boards. Run separate excitation and sense
wires for the Active, Reference, and Ground connections. The flow meter will not operate properly without these connections. If the Active
and Reference wires are reversed the flow meter will output incorrect flow rates. Use shielded cable with all signal applications. Connect
the shield to the input board only and not to the flow element terminal strip. Connect the shield to pin 1 of TS_ on the input board.
The connector for sensing point 1 must be plugged into the innermost connector on the input board which is plugged into P1 of the control
board. Sensing point 2 is connected to the outermost connector of the same input board. Sensing point 3 is plugged into the innermost
connector on the input board which is plugged into P2 of the control board. Sensing points 4 through 8 are plugged in the same manner.
Make the customer connections to the output modules (relay logic, current or voltage).
Summary of Contents for MT86
Page 1: ...Doc No 003162 Rev G...
Page 19: ...MT86 MT86HT Flow Meter OPERATION Fluid Components International LLC 15 Figure 9 A D Data Sheet...
Page 40: ...APPENDIX A DRAWINGS MT86 MT86HT Flow Meter 36 Fluid Components International LLC C01156 1 1...
Page 47: ...MT86 MT86HT Flow Meter APPENDIX C CUSTOMER SERVICE Fluid Components International LLC 43...
Page 48: ...APPENDIX C CUSTOMER SERVICE MT86 MT86HT Flow Meter 44 Fluid Components International LLC...
Page 54: ...MT86 MT86HT Flow Meter Fluid Components International LLC 003162 Rev G...