DEFIBRILLATOR MONITOR
Mod. 3850B-BIPHASIC - USER MANUAL
Page 48 of 63
9
ATTENTION! Do not store the device in deposits, or leave it between periods of use, in places where the sun shine directly
into the device. Risk of damage to the cover device, parts and accessories.
9
ATTENTION! During storage in warehouses and between uses, respects the conditions of temperature, pressure and
humidity as defined in this manual and periods of recharge the internal battery specified.
9
ATTENTION! The internal battery of this equipment cannot be replaced by the user. Necessarily must be replaced by a
qualified technician.
ATTENTION! In case you replace the battery follow the local instructions to dispose Ni-Mh batteries or
send them to
feas
ELECTRÓNICA
for disposal.
5. TROUBLESHOTING
GENERAL
Trouble Probable
cause
Action
Pushes the power on key, but the
equipment no power on.
The equipment is disconnected from
mains or +12Vext.
To connect firmly the cable to the
mains and to the equipment.
The display is dark; the waveforms
and characters are not sees.
The contrast is adjusted incorrectly.
Adjust the contrast as explained in
“Contrast”.
The date and/or time on screen and
on the paper record is not correct.
The date and/or time are incorrect.
Set Date and Time according “Date
and Time”.
No sync sound (QRS beep).
The motherboard is faulty.
Send the equipment to our
Customer Service for repair.
No audible alarm signal.
The
key (Alarm silence) is
activated.
Pres the
key (Alarm silence)
again.
The alarms are disabled “---” in all
alarm limits.
Select, on screen, the upper and
lower limits of each parameter
alarm.
The speaker, your cable or the
motherboard is faulty.
If the problem persists, the
motherboard is the cause, Send the
equipment to our Customer Service
for repair.
DESFIBRILADOR
Trouble Probable
cause
Action
The defibrillator doesn’t Charge
energy. The message “Charge Error”.
Operating from the battery, it is almost
empty.
Connect the equipment to an outlet
and recharge the battery. The battery
is charged automatically when the
equipment is connected to the mains.
Take more than 10 seconds to
complete the charge.
The defibrillator doesn’t Charge
energy. The message “Charge Error”.
Operating from the mains breakage in
the power system charge.
Send the equipment to our Customer
Service for repair.
Take more than 10 seconds to
complete the charge.
Pressing paddle buttons don’t
discharge energy to the patient.
If discharge is performed from “Shock”
key, then extern paddles are faulty.
Replace external paddles.
Pressing the “Shock” key, don’t
discharge energy to the patient.
Faulty keyboard.
Send the equipment to our Customer
Service for repair.
Pressing paddle buttons or discharge
front button, don’t discharge energy to
the patient.
Faulty equipment.
Send the equipment to our Customer
Service for repair.