10. Use of insoles with FeetMe® Evaluation mobile application

a. First use or reinstallation

 

Create a Gmail account if you do not have one. Go to the « Play Store » 
and search for the application : « Feetme Evaluation ». 

Download the application and follow the installation instructions.

b. Perform a gait recording

1. 

Connection to the insoles

 

Click on the button « Connect the insoles », select the right and left soles 
of the corresponding size and then click on « Next step ».

2. 

 Insoles calibration 

For this step, the insoles must be in the shoes, laid on a flat surface.  
Click on the « Calibration » button. 

3. 

Select a test

 

Click on the chosen test. 

You can also acquire gait parameters without the phone and synchronize 
the data at the end of the recording by clicking on « Import Recording ».

4. 

Start & stop recording

Put on shoes and press « Start » then « Play » to start recording. For  
« Standard » tests, press « Stop » and « Save » to save the data. For other 
tests, stopping the recording and saving the data are automatic.

5

.  Make a new recording

You remain connected by default to the last insoles, so you can make other 
recordings without redoing the steps of connection and calibration (repeat 
steps from step 3). To make a new recording with other insoles, click on  
« Connect the insoles » and repeat the above steps from the beginning.

c. View the data acquired from the mobile application

The data collected for a given patient can be viewed from the application. 
To access it, go to the « Recordings » tab to access all the records. To 
access the data of a record, click on the associated record.

d. Data backup

In case the user hosts the gait recordings, he is in charge of performing 
appropriate and regular data backups in order to prevent any destruction or 
loss of information. FeetMe cannot be held responsible for any accidental
or unauthorised destruction or accidental loss of user data.

e. View battery level

     

  

 c. Charge light indicators 

Continuous green : charged 
Continuous white :  charging
Discontinuous white : Insoles poorly positioned, reposition
Light off : power cable disconnected or improper positionning of 
the charger or the insole. Check position to make the light indicator 
steady white.

When the insoles are charging, the indicator light of the chargers 
is solid white. Once the insoles are charged, the indicator turns 
green. The indicator is flashing white when an insole is incorrectly 
positioned, the insole is not charging. It takes approximately 2.5 
hours for a complete charge of the insoles.

       

 d. Bluetooth connection to the insoles

Once charged, the insoles are automatically detectable via 
Bluetooth using FeetMe® Evaluation app. The insoles go to sleep 
mode automatically when not in use. In order to scan them, just 
shake them for 3 seconds. No manipulation is therefore necessary
from the user part, to turn the insoles on and off. If the insoles are
not detected, consider charging them. Be careful not to use the 
insoles alongside other Bluetooth devices.

Charging the insoles 
 

To position a pair of insoles 
in a charging slot, identify 
the insole size and locate the 
corresponding slot. Position 
each insole on the edge, top 
side facing outwards.

b. Charging the insoles with the charging case

Setting up the charging case

Open the case and remove the lid. The insoles wont charge if the lid 
is closed. Take out the power cable from the case and connect it  to  
the  case,  close  to  the  On/Off  switch.  Plug  the  other  end  to  a  
power  outlet.  Switch  it  on  (position  I).  If  the  charging  slots  are  
empty, the LED will flash white for a few seconds then go off.

The charging case should be at a temperature below 35° Celsius.

Summary of Contents for Monitor

Page 1: ...1 Instruction for use FeetMe Evaluation FeetMe Monitor GB FR...

Page 2: ...2 3 Instruction for use GB English...

Page 3: ...54 1 2A 5V RoHS LOT 0459 IP54 1 2A 5V RoHS LOT 0459 IP54 1 2A 5V RoHS LOT IP54 1 2A 5V RoHS LOT 1 2A 5V RoHS LOT 1 2A 5V LOT 1 2A 5V 0459 IP54 Table of content 1 Labels and symbols 2 Product scope 3 U...

Page 4: ...uation that allows visualization in real time of these parameters 3 User profile There are two types of users Health professionals who deliver insoles to their patients record and analyze the data Pat...

Page 5: ...30 cm 12 inches to any parts of the FeetMe Monitor insoles including cables specified by the manufacturer Otherwise the performance of these devices may be impaired Warning Use of accessories transdu...

Page 6: ...preventanydestructionor loss of information FeetMe cannot be held responsible for any accidental or unauthorised destruction or accidental loss of user data e View battery level c Charge light indicat...

Page 7: ...fall can alter the performance of the product Storage Warning In case of non use of your device it is advisable to keep it away from dust and in a clean and dry place FeetMe insoles should not be stor...

Page 8: ...stics Value Bluetooth v5 0 low energy Emitted power 4 dBm Frequency 2400 2483 5 Mhz Designation Value OS Android Version 5 1 1 Bluetooth v4 0 4 0 15 Technical characteristics Use condition of the inso...

Page 9: ...n and by the reflection of structures objects and people The field strengths of fixed transmitters such as base stations for radiotelephones mobile wireless and land mobile radios amateur radio AM and...

Page 10: ...18 19 Manuel d utilisation FR Fran ais...

Page 11: ...RoHS LOT IP54 1 2A 5V RoHS LOT 1 2A 5V RoHS LOT 1 2A 5V LOT 1 2A 5V 0459 IP54 Table des mati res 1 Explications des tiquettes et symboles 2 Domaine d application 3 Profil utilisateur 4 propos du pr s...

Page 12: ...er le dispositif et pr venir le fabricant ou le distributeur pour r paration ou remplacement Ne pas faire tomber vos semelles FeetMe Monitor ou leur faire subir de chocs importants Ne pas mettre le c...

Page 13: ...rs parce que cela peut occassionner un mauvais fonctionnement Si cette utilisation est n cessaire il convient d observer cet appareil et les autres appareils pour en v rifier le fonctionnement normal...

Page 14: ...gistrement Par d faut vous restez connect aux derni res semelles vous pouvez donc faire d autres enregistrements sans refaire les tapes de connection et calibration reprendre les tapes partir de l tap...

Page 15: ...er avec un linge propre sec et non peluchant D faut Cause Action entreprendre Semelle non d tectable Semelle d charg e Charger compl tement la semelles Bluetooth non activ Activer le Bluetooth de votr...

Page 16: ...s FeetMe Monitor Ne pas nettoyer les semelles FeetMe Monitor dans un lave vaisselle un s che linge ou tout autre appareil vapeur Ne pas immerger 14 limination Le dispositif contient des composants lec...

Page 17: ...res comme r orienter ou repostionner les semelles Dans la gamme de fr quence de 150 kHz 80 MHz il convient que les intensit s de champ soient inf rieures 3 V m Conformit FeetMe Monitor est conforme la...

Page 18: ...fabricant 19 Dur e de vie La dur e de vie du dispositif est la plus petite dur e entre deux ans ou unmillion depas faitpar l utilisateuravecleproduit soit500000pas par semelle Cette dur e ne s appliqu...

Page 19: ...36 FeetMe 157 boulevard MacDonald 75019 Paris France contact feetme fr Ref FTM DK AA IFU FR rev04 9 1 2A 5V 0459 IP54...

Reviews: