Operating Procedures
:
Specimen Preparation and Handling
C O N F I D E N T I A L – FEI Limited Rights Data
5-4
FIGURE 5-2
Power supply adaptor with removed plug / with mounted plug
Operation
The inner surface of the unit is polished and vacuum-clean. Avoid contamination of the inner surface otherwise the
specimens may become contaminated during the baking cycle.
C a u t i o n !
Always wear clean lint-free gloves when handling the unit and the specimens! Be sure the working desk is free from
impurities that can cause the unit or specimen contamination.
The working desk must be covered with a material with a melting temperature higher than 250 °C!
W A R N I N G !
Do not look directly at the operating bulb for any period of time! This may cause serious eye injury.
For support tray manipulation (sample loading for instance) always use the special tool delivered! Risk of burning!
Never touch the inner surface of the baking unit including the trays and specimens with bare hands or hands in gloves.
FIGURE 5-3
Support tray Handling tool / Removal of Support tray
1.
Open the lid and remove the support trays from the unit using the
special tool.
2.
Put the support tray on the working desk and put one or more
specimens on it so that the specimen stub pin enters the hole in the
tray.
3.
Put the tray back in the unit and close the lid.
There are two possible slots for the tray insertion. The specimen will
be exposed to the temperature of 160 °C in the upper position and of
120 °C in the bottom position.
Summary of Contents for Scios 2
Page 1: ...User Operation Manual Edition 1 Mar 2017 ...
Page 103: ...Alignments I Column Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 19 ...
Page 110: ...Alignments 254 GIS Alignment option C O N F I D E N T I A L FEI Limited Rights Data 4 26 ...
Page 170: ...Operating Procedures Patterning C O N F I D E N T I A L FEI Limited Rights Data 5 60 ...
Page 178: ...Maintenance Refilling Water Bottle C O N F I D E N T I A L FEI Limited Rights Data 6 8 ...