Operating Procedures:
Microscope Control
C O N F I D E N T I A L – FEI Limited Rights Data
5-7
Inserting / Exchanging Specimen
1.
Click on the
Vacuum
module /
Vent
button.
The confirmation dialog appears.
After switching a High Voltage off, the
vacuum system switches off the pumps and
opens appropriate valves to vent the system.
After a specified venting time the venting
valve will close.
Note
If the venting valve closes before the chamber is at
the atmospheric pressure (the door is not possible to
open), click on the Vent button once more to open it again.
2.
When vented, open the specimen chamber and, using lint-free gloves or tweezers, place a specimen into the
specimen holder.
3.
Close the specimen chamber door and click on the
Vacuum
module /
Pump
button.
Instructions for a sample exchange are also shown within the informational window by clicking on the
toolbar icon.
Selecting Vacuum Mode
When a specimen and appropriate detector(s) are installed
correctly, continue the procedure (see above):
4.
Check the
High Vacuum
module /
Low Vacuum
or
High Vacuum
mode radio button.
For the Low Vacuum mode select the appropriate gas from the
drop down list and the target
Chamber Pressure
that the system
pumps to.
W A R N I N G !
The system can be damaged by using the Low Vacuum mode without
an appropriate PLA. Do not select a PLA Cone which is not actually
used (see Chapter 2)!
X-Ray Cone Mounting / Dismounting Procedure
Under normal operation, the system will know the PLA cone size for
the detector installed in the LoVac mode. This will allow the system
to set automatic pressure range limits for the aperture installed,
thus avoiding vacuum errors when setting chamber pressure.
For the optional
GAD cone
– see Chapter 7.
The X-Ray cone is stored in the provided installation tool.
The large
arresting screw releases working part so it can slide along the skid.
A little spring is a conductive contact with the final lens pole.
Summary of Contents for Scios 2
Page 1: ...User Operation Manual Edition 1 Mar 2017 ...
Page 103: ...Alignments I Column Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 19 ...
Page 110: ...Alignments 254 GIS Alignment option C O N F I D E N T I A L FEI Limited Rights Data 4 26 ...
Page 170: ...Operating Procedures Patterning C O N F I D E N T I A L FEI Limited Rights Data 5 60 ...
Page 178: ...Maintenance Refilling Water Bottle C O N F I D E N T I A L FEI Limited Rights Data 6 8 ...