System Control:
Interface Elements
C O N F I D E N T I A L – FEI Limited Rights Data
2-3
Final Lens Aperture Strip
The strip is made from a Mo coated Si. The Automatic Aperture System (AAS) automatically controls the final lens
aperture selection according to imaging needs when working with the electron and ion beams (see Chapter 5).
Note
The aperture holes edges cleanness is very important, exchange or clean them if necessary (see Chapter 6). When
performing the E-Column: Aperture Selection alignment (see Chapter 4) there are knobs accessible on the aperture
mechanism on each column enabling the fine adjustment of the strip hole.
FIGURE 2-3
Automatic Aperture System Control Knobs Electron / Ion column
External Connectors panel
It is located at the back side of the microscope console.
It is used to connect 3rd party equipment to the
following connectors:
CONTROL SIGNALS / E-BEAM SCAN INPUT /
VIDEO OUT
These connectors are used for connection of EDX, WDX
and lithography systems, provided by 3rd party
suppliers. Do not connect anything else in these
connectors!
SPECIMEN CURRENT
This connector is used together with the Keithley
picoamper meter (option) equipment.
Ground point
The point is intended for a defined ground connection
between a microscope and a 3rd party module. This
connection is not a safe ground!
ELPHY SCAN & BLANKING IN
Not used.
Summary of Contents for Scios 2
Page 1: ...User Operation Manual Edition 1 Mar 2017 ...
Page 103: ...Alignments I Column Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 19 ...
Page 110: ...Alignments 254 GIS Alignment option C O N F I D E N T I A L FEI Limited Rights Data 4 26 ...
Page 170: ...Operating Procedures Patterning C O N F I D E N T I A L FEI Limited Rights Data 5 60 ...
Page 178: ...Maintenance Refilling Water Bottle C O N F I D E N T I A L FEI Limited Rights Data 6 8 ...