System Control
:
System States
C O N F I D E N T I A L – FEI Limited Rights Data
2-8
To bring the system back to an operation, follow the procedure:
C a u t i o n !
When the Standby state was just entered, wait 10 s at minimum before starting microscope operation again (pressing
the power button on the microscope front control panel)!
1.
Push the amber
Power
button on the Microscope console panel
– it changes to the green one
(a)
.
2.
Double-click on the
xT microscope Server
icon to start the
software (all seeming LED’s should be green – Initialized).
3.
Click on the
Start
button to start the server. Wait until all
elements are fully functional (green – Initialized).
4.
Click on the
Start UI
button to start the Microscope Control
software. The main window appears behind the
UI Log On
dialog.
Note
This step can take place automatically, if Advanced / Autorun UI
check box is ticked.
Now the system is in the Overnight state, to proceed to the
Full
operation
state follow the procedure:
5.
Enter your
Username
and a
Password
(b)
.
6.
Click on the
Column
module /
Beam On
button to start the
selected beam only or the
System
module /
Wake Up
(c)
button
to start both beams. A
Column
module source awaking
progress bar indicates the actual status and the button turns
orange when finished.
Note
d)
It is strongly recommended to always leave the chamber evacuated
when not being used. When the sample chamber is left in the LoVac
mode, water vapour is likely to accumulate in it, PVP lifetime
decreases and the water reservoir or gas cylinder empties
prematurely.
e)
Waiting for a new user leaves the status of the Microscope Control
software non-operational and only the xT microscope Server software
is active. Therefore changing a user does not require Logging off /
Logging on at Windows 7™ level.
Summary of Contents for Scios 2
Page 1: ...User Operation Manual Edition 1 Mar 2017 ...
Page 103: ...Alignments I Column Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 19 ...
Page 110: ...Alignments 254 GIS Alignment option C O N F I D E N T I A L FEI Limited Rights Data 4 26 ...
Page 170: ...Operating Procedures Patterning C O N F I D E N T I A L FEI Limited Rights Data 5 60 ...
Page 178: ...Maintenance Refilling Water Bottle C O N F I D E N T I A L FEI Limited Rights Data 6 8 ...