Alignments
:
C O N F I D E N T I A L – FEI Limited Rights Data
4-2
Alignments List
Grouped alignments lists are possible to roll down / up by
clicking the + / - sign:
Electron Beam
•
Magnification Correction
User magnification calibration (see Beam menu item).
•
E-Column: Aperture Selection
Enables to set a second set of final lens apertures in case the
image quality deteriorates (one or more actually used
apertures are worn out).
•
E-Column: Preventive Maintenance
The user level electron beam adjustment procedure.
•
E-Column: Supervisor Alignments
Set of Supervisor level electron beam adjustment procedures.
Ion Beam
•
100 – ION: Source Control
Used for ion source starting and heating, LMIS recovering,
LMIS state monitoring.
•
I-Column: Aperture Management
Controls the ion beam aperture strip holes condition.
•
I-Column: Manual Beam Alignment
•
I-Column: Alignments
Stages
•
Stage Alignments
– set of utilities:
– Stage Rotation Center
– sets the stage rotation centre for the compucentric rotation.
– Eucentric Height
– defines the eucentric position.
– R Zero Alignment
– sets zero rotation position of the holder for STEM range definition.
– Tilt Zero Alignment
– sets the stage tilt zero position.
– Tilt-X Correction
– sets the correction in X direction when tilting the stage.
EasyLift
•
EasyLift Needle Exchange/Calibration
(option – see Chapter 7)
Others
•
154 – Water Bottle Venting
Enables to vent the water bottle in case of refilling.
•
External Plasma Cleaning
/
Plasma Cleaning
Enables the chamber plasma cleaning and sets duration of the specimen plasma cleaning.
•
Pump Buffer
Enables to pump down the vacuum buffer.
•
Vacuum Actions
(U) – set of utilities:
– Start / Stop IGPs
– enables to start / stop vacuum pumps in case of power failure, emergency or service actions
– Pump / Vent Actions
(U) – enables to pump / vent particular vacuum system sections
•
GIS Alignment
(option)
Resets the GIS crucible filling lifetime counter after its exchange, sets the GIS heating temperature and performs
GIS mechanical alignment.
Model difference
Scios 2 High Vacuum has a slightly modified alignments list.
Summary of Contents for Scios 2
Page 1: ...User Operation Manual Edition 1 Mar 2017 ...
Page 103: ...Alignments I Column Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 19 ...
Page 110: ...Alignments 254 GIS Alignment option C O N F I D E N T I A L FEI Limited Rights Data 4 26 ...
Page 170: ...Operating Procedures Patterning C O N F I D E N T I A L FEI Limited Rights Data 5 60 ...
Page 178: ...Maintenance Refilling Water Bottle C O N F I D E N T I A L FEI Limited Rights Data 6 8 ...