Alignments:
Automatic Alignments
C O N F I D E N T I A L – FEI Limited Rights Data
4-3
Automatic Alignments
Some of Alignments (or a set of alignments) have a different
approach and most of the procedures can run automatically by
checking the
Automatic
check box placed at the bottom of the
alignment module. Automatic progress can be started / stopped
at any time during the procedure. Accessible informational
windows can be turned on / off by the
Show
/
Hide Window
button. The alignment sequence can be cancelled
/ continued by the Cancel
/ Next
button.
C a u t i o n !
Read carefully the instructions provided by the system before proceeding with any alignment!
If you have any e-beam alignment procedure selected and you want to select
any other one, you must
Finish
or
Cancel
an actually opened procedure first.
To start alignment procedure or to stop one in progress right-click on the red
circle and select required action (
Start
/
Stop
– variable) from the context
menu. A confirmation dialog is shown.
When interrupting any alignment
(clicking on the
Cancel
button, selecting the
Stop
item in the menu, closing the UI)
nothing is saved. In all cases the alignment
windows disappear and the
Stop Execution
dialog appears.
The last page of every alignment offers one
or more of following choices:
•
Restore old values and finish
: undoes the
adjustments, nothing will be saved.
•
Save new results and finish
: keeps new
adjustment values and saves them.
Summary of Contents for Scios 2
Page 1: ...User Operation Manual Edition 1 Mar 2017 ...
Page 103: ...Alignments I Column Alignments C O N F I D E N T I A L FEI Limited Rights Data 4 19 ...
Page 110: ...Alignments 254 GIS Alignment option C O N F I D E N T I A L FEI Limited Rights Data 4 26 ...
Page 170: ...Operating Procedures Patterning C O N F I D E N T I A L FEI Limited Rights Data 5 60 ...
Page 178: ...Maintenance Refilling Water Bottle C O N F I D E N T I A L FEI Limited Rights Data 6 8 ...