up   and   make   sure   it   springs   back   to   the   center.   Move   the   storage   block   
back   down   again   to   the   cartridges   remain   submerged.   Top   up   the   LN2.   

30.

 

Repeat   for   all   other   positions.   If   a   cartridge   is   present   on   the   holder   
(Cartridge   button   Yellow)   allow   one   free   position   to   unload   it.   

   

   
   

   
Transferring   the   Selection   Rod   to   the   Microscope   

   

31.

 

Once   all   positions   are   loaded   make   sure   V2   and   V3   are   closed   and   turn   on   
the   scroll   pump.   

32.

 

Open   V3   and   wait   for   the   pressure   to   fall   to   below   8x10

-­‐2   

Torr.   

33.

 

Retract   the   storage   block   inside   the   Selection   Rod   and   close   VC.   

34.

 

Open   V2   and   wait   for   the   pressure   to   fall   to   below   8x10

-­‐2   

Torr.   

35.

 

On   the   microscope   remove   the   airlock   cover   by   unclamping   the   two   
clamps   and   wipe   the   0-­‐ring   with   lens   paper.   

36.

 

Close   VA.   

37.

 

Close   V3.   

38.

 

Open   V4   to   vent   and   bring   the   interface   between   VC   and   VA   to   

atmosphere.   

39.

 

Unclamp   the   Selection   Rod   and   swing   to   the   right   to   undock   from   the   
serial   port.   Lift   up   to   un-­‐hinge.   

40.

 

Attach   the   selection   Rod   to   the   airlock   hinges   and   swing   in   to   dock.   Secure   
the   top   clamp   first   followed   by   the   bottom   clamp.   

41.

 

Clock   on   Pump   Airlock.   

42.

 

V43   will   open   followed   by   V41.   Once   P10   gets   to   Log   36   the   message   to   
open   VA   will   appear.   

43.

 

Pull   VA   to   open   and   wait   for   the   pins   to   pop   in.   

44.

 

A   prompt   will   shortly   follow   asking   to   open   VB.   

45.

 

Pull   VB   to   open   and   wait   for   the   pins   to   pop   in.   The   Selection   Rod   is   now   

connected   to   the   Microscope   vacuum.   Turn   the   Turbo   off   and   wait   for   the   
Column   vacuum   to   fall   below   Log   15.   

   

Removing   a   cartridge   from   the   holder

   

   

46.

 

Move   the   insertion   Rod   to   the   first   stop,   the   checking   position   and   check   
in   the   airlock   mirror   that   a   cartridge   is   not   present.   If   a   cartridge   is   

present   follow   steps   50-­‐53.   

47.

 

Retract   the   Insertion   Rod   and   click   on   Cartridge.   This   will   move   the   stage   
to   the   default   exchange   XYZ   coordinates   and   unlock   the   solenoid   safety   

pin.   

48.

 

Move   the   Insertion   Rod   slowly   inwards   to   the   final   stop   and   ensure   the   
silver   button   pops   up.   Once   locked   double   check   by   gently   trying   to   move   

the   insertion   rod   to   ensure   it   is   locked   in.   

49.

 

Unscrew   the   specimen   holder   from   the   cartridge   by   turning   the   white   

knob   anti-­‐clockwise   until   the   red   band   is   clearly   visible.   The   Cartridge   
button   should   now   be   grey.   

Summary of Contents for Tecnai F30 Polara

Page 1: ...ur and should not go below 20 5 Use the metal funnel to top up the Insertion Rod dewar This dewar can last up to 10 hours If the Insertion Rod warms up the vacuum deteriorates and the HT may switch off When collecting with EPU a LN2 top up at 11pm will ensure it stays cold until 9am the next day 6 Check the temperatures of the different channels on the display 1 Helium cryo shield 2 Nitrogen cryo ...

Page 2: ...ition should be accessible 18 Remove old grids and C clips from the cartridges using 5 tweezers 19 Record the bottom and top cartridge identifier digits on the notepad 20 Begin cooling the transfer chamber initially without the lid and then through the side funnel to avoid condensation inside Continue cooling the chamber and keeping the LN2 level just above the cartridge loading area 21 After a co...

Page 3: ... to dock Secure the top clamp first followed by the bottom clamp 41 Clock on Pump Airlock 42 V43 will open followed by V41 Once P10 gets to Log 36 the message to open VA will appear 43 Pull VA to open and wait for the pins to pop in 44 A prompt will shortly follow asking to open VB 45 Pull VB to open and wait for the pins to pop in The Selection Rod is now connected to the Microscope vacuum Turn t...

Page 4: ...rds to the final stop and ensure the silver button pops up Once locked double check by gently trying to move the insertion rod to ensure it is locked in 61 On the holder push in the white knob until the red stripe disappears and then screw clockwise one complete turn to capture the threads on the cartridge Continue turning until finger tight DO NOT OVERTIGHTEN The Cartridge button should turn Yell...

Page 5: ... ...

Page 6: ...Spirits one session is up to half a day 4 hours for negative stain and up to a full day 8 hours for cryo EM A half day is for example 9am 1pm 1pm 5pm or 5pm 9pm A whole day would be 9am 5pm You can of course book less time if you don t need the full 4 or 8 hour slot The actual booking times are also flexible e g 10 12pm Please note that evenings and weekends count the same on the Spirits If on the...

Reviews: