FZ12 Microscope Operation:
13
Fein
Selecting the Light Path:
When the beam splitter pole (1) is in the
innermost position, the light is directed to
the binocular head for observation through
the eyepieces. When the beam splitter pole
is in the outermost position, the light is
directed to the trinocular tube which is used
for observation using a camera.
Assembling the C-Mount Adapter:
Loosen the lock screw (1) on the trinocular tube
and remove the dust cap (2). Remove the two
dust caps (3) on the c-mount adapter. Insert the
c-mount adapter into the trinocular tube and
secure in place with the lock screw (1). Connect
the camera to the c-mount adapter. Ensure
the beam splitter pole (4) is in the outermost
position for observation using the trinocular
tube. If the image is not clear, loosen the lock
screw (5) on the c-mount adapter and rotate
the focusing portion of the c-mount (6) until the
image is clear. Once the image is clear, tighten
the lock screw (5).
Assemble and Use the Polarizer Device:
Remove the stage plate (2) from the base
and insert the polarizer (1) onto the base.
Attach the analyzer (3) to the outside
diameter of the main objective (4) and
secure it in place with the lock screw. The
analyzer (5) is adjustable by 360°. Rotate the
analyzer for adjustment.
1
1
2
3
5
6
4
1
3
2
4
5