Bulb Replacement:
13
Fein
Before performing any of these steps the
microscope MUST be turned off and unplugged.
With the microscope powered off and
unplugged, gently lay the microscope down on a
sturdy surface ensuring the carrying handle side
is down.
Locate the 4 rubber feet on the underside
of the microscope. Remove the rubber feet
exposing the 4 screws securing the bottom
plate of the microscope.
Gently lay microscope on
carrying handle.
Using an Allen wrench remove the 4 screws
securing the bottom plate of the microscope.
With all 4 screws removed, gently remove the
bottom plate exposing the internal electrical
components of the microscope.