Step-by-Step Assembly:

info@feinoptic.com

Page 10

Replacing the Fuse:

Turn the power switch to “O” off. 

Remove the power cord.

Unscrew the fuse group (1) from the base (2) with a 

flathead screwdriver. 

Install a T250V, 3.15A fuse.

C-Mount Assembly:

Loosen the trinocular head set screw (1) and remove 

the trinocular port dust cap (2).

Remove any dust cover caps from the c-mount (3).

Insert the c-mount into the trinocular port as image 

shows at right and tighten the set screw (1).

Loosen the c-mount set screw (4).

Connect the camera to the c-mount threads (5) and 

tighten the set screw.

When the eyepieces are in focus, adjust the focus on 

the c-mount with an allen key to parfocal the camera 

with the eyepieces.

Phase Contrast Slider Assembly:

Insert the slider (1) from left to right with the 

inscription facing up.

Each hole in the phase slider has a corresponding 

position. 

When using phase contrast make sure the 

adjustment ring on the aperture diaphragm (2) is set 

to “PH”.

Each phase objective should be used with the 

corresponding phase ring diaphragm (for example 

10xPH with 10x ring).

The phase slider is pre-centered.

Darkfield Slider Assembly:

Insert the slider (1) from left to right with the 

inscription facing up.

When using darkfield adjust the aperture diaphragm 

to an open position.

The darkfield ring diaphragm adjustment (1) is 

located on the left side of the slider.

Summary of Contents for RB30

Page 1: ...RB30 Microscope User s Manual info feinoptic com...

Page 2: ...info feinoptic com Page 1 Microscope Components...

Page 3: ...igh temperatures dust or damp environments Use a flat work surface Set the power switch to off O before replacing a bulb or fuse and wait until the lamp is cool The power switch 4 is located on the si...

Page 4: ...he detergent is removed from the frame with a clean damp cloth prior to drying the surface Do not disassemble the microscope After use cover the microscope with a dust cover and power off the light Ob...

Page 5: ...Binocular Trinocular Microscope Assembly info feinoptic com Page 4...

Page 6: ...Fluorescence Microscope Assembly info feinoptic com Page 5...

Page 7: ...operating the microscope always start using the lowest objective 4x or 10x to get the sample into focus before moving to a higher magnification Eyepiece Assembly Remove the cover from the eyetube 1 In...

Page 8: ...e brightness and counterclockwise to decrease brightness Using the bulb at a lower brightness extends bulb life Slide Placement Push the specimen holder 1 back Place the slide 2 in the slot with the c...

Page 9: ...until the image is clear If multiple people are using the microscope note your diopter setting for quick transition from users Interpupillary Distance Adjustment When observing with both eyes adjust t...

Page 10: ...p with the inscription 4 of the objective being used Using the 100x Immersion Oil Objective Start with the 4x objective and focus on the sample Place a drop of immersion oil 1 on the specimen Rotate t...

Page 11: ...eyepieces are in focus adjust the focus on the c mount with an allen key to parfocal the camera with the eyepieces Phase Contrast Slider Assembly Insert the slider 1 from left to right with the inscri...

Page 12: ...move one eyepiece and insert the phase centering telescope Set the phase adjustment ring 2 to correspond with the phase objective being used example PH10x with 10x objective Adjust the centering teles...

Page 13: ...the binocular trinocular head on top of the fluorescence module and tighten the set screw 3 Insert transformer 1 into socket 2 and then plug directly into outlet Fluorescence Light Adjustment Fluores...

Page 14: ...r glass thickness 0 17mm Cover slip is on the bottom of the slide Put cover glass face up Immersion oil has dried on objective lens Clean objective lens No oil immersion was used with 100x lens Use im...

Page 15: ...ched properly Adjust specimen holder Stage is not fastened properly Reattach the stage Power is not turned on Check power cable connection LED light is not inserted properly Check light connection LED...

Page 16: ...to each 3 Disconnect the wires from the bulb assembly and unscrew the bulb assembly from the bulb compartment 4 Solder the wires to the new bulb assembly in the same location as the old bulb assembly...

Reviews: