Step-by-Step Assembly:
info@feinoptic.com
Page 6
Condenser Assembly:
Rotate the coarse focus knob (1) to raise the stage to
its highest position.
Rotate the condenser lowering knob (2) to lower the
condenser bracket.
Loosen the condenser lock screw (3).
Insert the condenser and tighten the lock screw.
Raise the condenser back to the highest position.
Objective Assembly:
Rotate the coarse focus knob (1) to lower the stage.
Install the objectives (2) into the nosepiece (3) from
the lowest magnification to the highest in a
clockwise direction from the rear.
When operating the microscope always start using
the lowest objective (4x or 10x) to get the sample
into focus before moving to a higher magnification.
Eyepiece Assembly:
Remove the cover from the eyetube (1).
Insert the eyepiece (2) into the eyetube.
When adjusting the diopter on the eyepiece, ensure
the eyepiece is locked into the eyetube with the hex
screw (3) to avoid having the eyepiece rotate in the
eyetube.
Light Source Assembly:
Align the orientation pin (1) and the power pin (2) to
the holder (3) and socket (4).
Push the light source into the arm smoothly and
plug it in.
Whenever replacing the bulb, turn off the power and
ensure the bulb holder is not warm.
Replace entire LED light source if LED bulb is burned
out. See page 15 of this manual for instructions.