F-960 Instruction Manual rev. 6/23/2017
1554 NE 3
rd
Ave, Camas, WA 98607, USA
3
Phone: (360) 833-8835
sales@felixinstruments.com
www.felixinstruments.com
F-960 Specifications
Air Sampling Rate
70 mL/min
Measuring Rate
Automated, 1 second intervals to SD
in Continuous Mode
Data Storage
Removable 16 GB SD and Bluetooth
Display
Sunlight visible transflective LCD
Operating environment
0°C - 45°C (0-90% relative humidity,
non-condensing)
Power Source
Removable rechargeable lithium-ion
battery
Dimensions
18cm x 13.5cm x 5.5cm
Weight
0.95kg
Enclosure
Powder coated aluminum
Warm Up Time
<3 minutes
Sensors
Ethylene (C2H4)
SENSOR
Electrochemical
Nominal Range
0-1000 ppm
Full Scale Resolution
1 ppm
Accuracy Trigger Mode
±5% relative 10 ppm absolute
Accuracy Continuous
Mode
±5% relative 2 ppm absolute
Lower Detection Limit
10 ppm
Sampling Time
35 seconds
Zero Interval
Weekly
Recalibration Interval
3 months
Lifetime
12 months