7
Three Intensity Levels
The PowerStrobe provides three intensity levels for tailoring the flashlamp
brightness to the actual inspection application.
The flashlamp intensity is set by using the intensity control key to
select 100%, 66%, and 33% levels, depending on the operating flash rate.
The selected intensity level also selects the flash rate range and auto-
matically limits the maximum flash rate.
Three Flash Rate Ranges
Maximum flash rates of 6000 F/M, 12000 F/M, and 18000 F/M are
provided and are automatically set by the selected intensity level as fol-
lows:
Intensity Level Flash Rate Range
100% 30 F/M to 6000 F/M
66% 30 F/M to 12000 F/M
33% 30 F/M to 18000 F/M
Built-In Memory Feature
The PowerStrobe contains a built-in memory feature that automatically
updates and stores its last used control setup, each time the push but- ton
on/off trigger switch is released.
Control parameters are automatically stored as long as they remain unchanged
for a 5 second period before releasing the on/off trigger switch.