© Ferno / 234-1929-06 / July 2021
18
Model 35-ITA Incubator Transporter
6.3 Cleaning the Transporter
1. Hand clean all surfaces of the transporter with warm water and a
mild detergent.
2. Rinse with warm, clear water. Dry the transporter with a towel or
allow it to air-dry.
To clean the accessories, follow the cleaning instructions in their users’
manual.
6.4 Waxing the Transporter
While it is not necessary to wax this transporter, waxing the aluminum
main frame or the painted legs will not damage them, and will help
maintain the transporter’s appearance.
Disinfect and clean the transporter before applying wax. Use an
automotive wax as directed by the wax manufacturer.
6 - MAINTENANCE
6.1 Maintenance Schedule
The transporter requires regular maintenance. Set up and follow a
maintenance schedule. The table at right represents minimum intervals
for maintenance.
Keep maintenance records. For a sample record sheet, see “Maintenance
Record” on page 23.
When using maintenance products, follow the manufacturer’s directions
and read the manufacturer’s material safety data sheets.
6.2 Disinfecting the Transporter
Wipe all surfaces with disinfectant. Follow the disinfectant manufacturer’s
instructions for application method and contact time. Ferno recommends
you inspect the transporter for damage as you disinfect it.
Minimum
Maintenance
Each Use
As Needed
Monthly
Disinfecting
(this page)
•
Cleaning
(this page)
•
Waxing
(this page)
•
•
Inspecting
(page 19)
•
Lubricating
(page 19)
•
NOTICE
Disinfectants and cleaners containing bleach, phenolics, or
iodines can cause damage. Do not use products containing
these chemicals.
Water under high pressure, or steam, can penetrate joints,
flush away lubricant, and cause corrosion. Use caution when
cleaning moving parts such as joints and hinges, and reapply
lubricant if needed. See “Lubricating the Transporter” on
page 21.
Using abrasive cleaning compounds or applicators on the
transporter can cause damage. Do not use abrasive materials
to clean the transporter.
WARNING
Improper maintenance can cause injury. Maintain the
transporter only as described in this manual.