8. Peak Hold Mode
20
Under this mode, the instrument will display and hold the highest load/force
detected.
1.
Refer to
5.5
on how to select the desired weight unit.
2.
Refer to
5.6
on how to enter Peak Hold mode.
3.
Complete all necessary test setup. If mounting/support accessories are
used, apply all of them.
4.
Press
[Tare]
to cancel the effect of any extra loads.
5.
Start measuring process, the peak value detected will be held and
flashing.
6.
To refresh / display actual current value (e.g. after a tension force has
been decreased), press
[Hold]
again.
9. Animal Weighing Mode
21
9.1 Description of Animal Weighing Mode
Animal weighing mode is used to weigh live animals.
9.2 Basic Animal Weighing Settings
1.
Refer to
5.5
on how to select the desired weight unit.
2.
Enter Animal Weighing mode.
3.
Press and hold
[Set]
for 2 second to enter weight release parameter
setting.
4.
Press
[Func] or [Unit]
key to select the preferred weight release
variation value. 10 parameters are available from Off to 20): -
5.
10 parameters are available from Off to 20): -
•
rE oFF = auto release disabled.
•
rE 0.5 = auto release when weight varies ≥0.5% of rate.
•
rE 1 = auto release when weight varies ≥1% of rate.
•
……….
20
Peak hold mode does not support memory accumulation, weight unit conversation or
weight check function.
21
Animal Weighing function will not operate when weight is less than 20d (or 20d
1
for dual
weighing range/interval).
Summary of Contents for CSa
Page 2: ...2 ...
Page 9: ...9 2 Specifications ...
Page 10: ...10 3 Keys Display Connection Points ...
Page 53: ...53 Appendix B PC Output Protocols ...
Page 54: ...54 ...
Page 55: ...55 ...
Page 56: ...56 ...
Page 72: ...72 Fidelity Measurement Co Ltd www fi measurement com e mail info fi measurement com ...