16. Built-in Battery & Recharging
16.1 Battery Operation Time
Depends on the battery operation condition, a new and fully charged
rechargeable battery can provide 40~60 operation hours depend of actual
power consumption. The following settings help saving extra energy: -
•
F8 Brightness: - the brighter the LED indicator, the shorter the
battery service time.
•
F16 and F17 Comport: turn off any comport which is not used.
•
F31 Auto Power Saving: - always set this to On. This will help save
power especially when the instrument is unattended.
•
F38 WIFI and Bluetooth: - turn off any device which is not used.
16.2 Battery Low Signal
When battery level is low, bAt.Lo alter will appear. Apply the external charger
or replace the battery with a fully charged one immediately.
To protect the built-in rechargeable battery, this instrument will be powered off
automatically when battery is at extremely low level. If this is the case, do not
attempt to power this instrument on. Fail to do so may cause unrecoverable
damages to the built-in rechargeable battery.
16.3 Battery Charging Status
Battery charging status is shown on the dual color
Charge Status Indicator
on
the external battery charger: -
•
Red: - Recharging in process,
•
Green: - Charging completed.