Hydro Logger H40
User manual
FIEDLER
electronics for ecology
39
5.2.
Replacing the battery power pack
Control channel K11 displays the remaining capacity of the battery pack, expressed as
a percentage. If this value falls below 15%, it is advisable to replace the power pack with
a new one. The power pack can be replaced by the user or ordered from the station
manufacturer. The power supply unit is equipped with a short cable ending with a con-
nector for easy connection to the electronic board of the instrument.
The Hydro-Logger H40G is powered by a lithium battery pack with a voltage of 3.6 V
and a capacity of 26 Ah.
The Hydro-Logger H40D without GSM communication module is powered by a smaller
3.6V / 19Ah block.
When replacing the battery pack, proceed as follows:
•
Unscrew the locking screw that secures the front panel in position in the Hydro
Logger H40's stainless steel cylindrical housing.
•
Slide the electronic part of the instrument connected to the front panel out of
the stainless steel housing of the instrument.
•
Disconnect the power connector from the electronics, remove the old battery
pack from the instrument case and replace it with a new one, connect the
power connector to the electronics part and insert it back into the instrument
case.
•
Apply silicone petroleum jelly to the O-ring in the front panel to prevent mois-
ture from entering the station and slide the front panel with the connected elec-
tronics back into the stainless steel housing, securing it with the locking screw.
•
Using the connected PC and the MOST program, RESET the capacity coun-
ter.
After replacing the power supply battery with a new one, you need to reset the power
draw counter using the MOST program and set the initial capacity to 100%.
To reset the counter, press the
Reset
button in the "
Main Information"
window. This win-
dow is called up by selecting
Info
->
Current values
from the main menu and selecting
the
Instrument status
tab.
Notice
The used power battery blocks can be returned to the manufacturer of the device
(FIEDLER AMS s.r.o. ,
Lipová 1789/9
, 370 05
České Budějovice), who has a contract
for the take-back of used batteries. Improper disposal of used batteries could damage
the environment.
H 4 0 G
H 4 0 D
S e t t i n g t h e i n i t i a l
c a p a c i t y a f t e r
b a t t e r y r e p l a c e m e n t
Button for zeroing
counters of the energy
consumed so far from
the power batteries
Press the button to set
the control channel K11
to 100% after inserting
new batteries.