background image

• With the drive box seated horizontally, secure

using the hex nut.

• Place the microscope head back into the drive

box and seat securely using the clamp screw (3).

Joint adjustment

The joint arm has a friction joint that is addi-

tionally supported at the centre and in the foot by
internal springs.
In order to adjust the joint arm, the wing nuts on the
joint arm must not be opened. The wing nuts are
used for basic adjustment of the friction and for
occasional readjustment.

Note:

The foot joint is adjusted using the quick-

release lever. The quick-release lever can transmit
forces that are so strong that the joint is completely
blocked. In this case, it is essential that the quick-
release lever is opened slightly prior to adjusting the
joint arm, as the joint arm will otherwise be dama-
ged.

Tip:

If the lever is pulled a little to the side away

from the joint, the lever can be moved to any requi-
red position without opening the joint.

Using the microscope:

• Place both ocular viewfinders (8) on the oculars

(1).

• Unscrew the protective cap from the objective

tube (4) by turning it anticlockwise.

• Look through both oculars (1) and adjust the ocu-

lar distance in line with the interpupillary distance
by turning the objectives (9) inwards or outwards.

• Place the specimen to be viewed under the

microscope on the specimen stage.

• In the case of larger specimens, the distance 

setting selected using the drive knobs at the side
may not be sufficient.

Adjusting the distance using the joint arm (see
sec tion on joint adjustment).

Guide values:

W= Widefield; Oculars 10x (supplied), other magnifi-
cations optionally available.

• Required magnification (1x or 2x) by turning the

objective tube (4) by 90° in each case.

• For fine adjustment, use the drive button (5) to

modify the distance to the object in order to
achieve maximum image precision.

• Set dioptre compensation (2) to suit individual

requirements.

• The working distance and field of vision can be

further increased using the optional FINO add-on
lens with 0.5x reducing optics (item no. 86006).
The add-on lens can be screwed onto the 
quick-action revolver.

Objective

magnification

Ocular

magnification

Overall 

magnification

Working

distance

(mm)

Field of view

ø (mm)

Field number =

Field of view ø 

(mm) x objective

magnification

1x

W 5x

5x

79

20

20

2x

W 5x

10x

84

10

20

1x

W 10x

10x

83

19

19

2x

W 10x

20x

83

9,5

19

1x

W 15x

15x

83

12,4

12,4

2x

W 15x

30x

83

6,2

12,4

1x

W 20x

20x

83

9,6

9,6

2x

W 20x

40x

82

4,8

9,6

6

Summary of Contents for 86000

Page 1: ...a n i a FINO der feine Unterschied the fine difference la fine différence la fina diferencia la fine differenza 86000 86001 mit Stativ mit Gelenkarm with table stand with articulated arm avec trépied de table avec un bras articulé con el pie de sobremesa con el brazo articulado con stativo con braccio snodato met statief met scharnierarm se stativem s kloubovým ramenem med stativ med ledad arm áll...

Page 2: ... setup 3 3 Safety precautions 4 4 Getting started 4 Prior to commissioning location 5 Handling 4 Using with a stand item no 86000 Using with a joint arm item no 86001 Working with an LED ring light 6 Service and maintenance 7 7 Storage 7 8 Technical data 7 9 Delivery forms 7 10 Guarantee 7 11 Disposal 8 CONTENTS 2 ...

Page 3: ... connect the episcopic illumination ring light of the device to a properly connected mains socket with a mains voltage of 230 V 50 Hz using the mains adapter supplied Check the mains adapter power cable and device for damage prior to operation If damage is identified take the device out of service and ensure that it cannot be put back into operation Have the device repaired by an authorised specia...

Page 4: ...atures direct sunlight strong vibra tions blows or impacts 5 Handling 5 1 Using with a stand item no 86000 The device is supplied almost entirely pre assem bled Raise the microscope knob a few centimetres using the drive knob 5 Loosen the clamping screw for the microscope head 3 and turn the microscope knob so that the oculars 1 are facing the user Secure the clamping screw 3 again Place both ocul...

Page 5: ...the swivel move ment of the joint arm Flexible assembly of the joint arm using the table clamp The work surface is not damaged Unscrew the four pre assembled machine screws from the table clamp supplied For easier assembly remove the screw on flange from the lower end of the joint arm To do so unscrew and remove both socket head screws from the side of the flange using the socket wrench supplied a...

Page 6: ...s 1 and adjust the ocu lar distance in line with the interpupillary distance by turning the objectives 9 inwards or outwards Place the specimen to be viewed under the microscope on the specimen stage In the case of larger specimens the distance setting selected using the drive knobs at the side may not be sufficient Adjusting the distance using the joint arm see section on joint adjustment Guide v...

Page 7: ... Store unit in a dry and dust free place 8 Technical data 8 1 With stand item no 86000 Metal stand column length 18 0 cm Flat base 20 0 x 15 0 cm Widefield oculars pair WF 10x Revolver 1x and 2x for magnification 10x 20x LED episcopic illu mination ring light 100 240 V 50 60 Hz 3 6 W Power cable 2 m 8 2 With joint arm item no 86001 Metal joint arm 85 0 cm Widefield oculars pair 10x Magnification 1...

Page 8: ...t time and then use it FINO GmbH cannot be held liable for damages due to improper use or due to modifications made to the device itself The warranty will be voided if the unit is used for any purpose other than its proper intended purpose FINO GmbH has a policy of conti nuously developing and improving its products FINO GmbH reserves the right to make changes or impro vements to the products desc...

Page 9: ...9 ENU ...

Page 10: ...FINO GmbH Mangelsfeld 18 D 97708 Bad Bocklet Tel 49 97 08 90 94 20 Fax 49 97 08 90 94 21 info fino com www fino com ...

Reviews: