Do not position material in contact with or near the temperature probes (Figure 14).
Figure 14
–
probes position
✓
Do not block air vents (Figure 15) and pressure equalising valve in SUPERARTIC models (Figure 16).
Figure 15
–
air vents at the top of the compartment
Figure 16
–
position of the SUPERARTIC model pressure equalising valve
3.9
INDICATION FOR OPTIMAL USE
The following instructions are provided to the user to follow in order to get the best equipment performances:
-
The power supply must comply with the information provided on the technical data plate (+/- 10%).
-
The units have been designed and built to work in environment with temperatures falling within the climatic class
temperatures indicated on the technical data plate (see Par. 9.1) and at a relative humidity of 60%.
-
Do not block the motor compartment air vents.
-
Load stored material gradually at ambient temperature to grant proper refrigeration.
-
Store material on shelves (or drawers). Do not place products directly on the bottom, or against the wall, doors or
fixed guards of the unit (see Par. 4.6.3)
-
Make sure doors are closed properly.
-
Limit door opening frequency and duration. Each time the door is opened, the internal temperature will alter and
there will be possible ice formation on the evaporator
-
Keep the defrost water drain outlet clear.
-
Follow a regular maintenance schedule (see Par. 6)