MNL004_EN rev. F
Pag.
30
di
56
4.3.3.3
How to change temperature alarm limits
Press the button
, then use
or till the following string
.
Confirm with the button
. The pre-set temperature alarm limits will be visualized and the lower limit will be
blinking too. Change it using the buttons
and
, then confirm by pressing the button
. After the
lower limit modification, the high temperature alarm limit will start blinking: use the buttons
and
to
modify it and confirm
.
Sequence with one probe (a):
Sequence with two probes (a
–
b):
To instantly visualize the pre-set limits press for some seconds the button .
For the models equipped with the ballasted product simulation temperature probe (standard only for
Medical devices)
: for single temperature models equipped with no. 2 probes in the chamber,
Sa
refers to the
product temperature simulation probe, whereas
Sb
refers to the air temperature probe.
4.3.3.4
Door openings visualization
Press the button
, then use
or till the following string
. Confirm with
the button
. The display visualizes the last recorded door opening:
LIMITS m2
DOOR OPENINGS
m4
Na = 02/12
d=004’
La=-20°
Ha=-10°
Lb=+02°
Hb=+08°
<<CONFIRMED>
La=-20°
Ha=-10°
>>CONFIRMED<
Door openings
number
(Chamber
A=Na)
(Chamber
B=Nb)
Number of
critical door
openings
(over 30 sec.)
Total number
of door
openings
Door openings
total daily time
in minutes