VISUAL AND ACOUSTIC ALARM
SOLUTION
COMPRES WORK.
•
Air condition the environment.
•
Clean the condenser (see par. 6.3)
•
Check that installation has been properly performed (see par. 4.2)
•
Switch-off the equipment from keypad (see par. 5.2.3.1) and then,
switch it on again.
•
Check that door is properly closed.
DEFROST TIME
•
Switch-off the equipment from keypad (see par. 5.2.3.1) and then,
switch it on again.
•
Check ambient conditions (environment should be neither too hot
nor too cold).
•
Check that installation has been properly performed (see par. 4.2)
•
If the problem persists, contact Technical Support.
HIGH TEMP alarm
•
Check that products have been correctly stored in the chamber (see
par. 4.6.3)
•
Check ambient conditions (environment should be neither too hot
nor too cold).
•
Clean the condenser (see par. 6.3)
•
Check that door is properly closed.
•
Air condition the environment.
LOW TEMP alarm
•
Contact Technical Support.
POWER FAIL HT
•
Check the alarms list.
•
Check the electrical system or accidental disconnection of the plug.
H 00:m00
•
Replace button battery on electronic board.
NO BATTERY (only if DMLP
Digital Monitor is installed)
•
Replace the battery.
BATTERY FAILURE (only if DMLP
Digital Monitor is installed)
•
Replace the battery.
PROBE S1-S2-S3-S4
•
Contact Technical Support. Refrigerator functioning is guaranteed
anyway.