PROTECTION
SYSTEM
Turn the refrigerator off from
the display, -STAND BY-
Par.5.2.8. After a while turn it
on and the visual and
acoustic alarm is off.
Tap the off alarm/fault icon so
to permanently delete the
signal (par.5.2.16).
Faulty electrical parts
The device is in a condition of
forced stop due to a failure that
potentially could be dangerous for
the integrity of other components
Display shows “WAITING” in the
“mode in progress”, Par.5.2.2
If temperature does not lower and it does not reach set
point, TURN OFF the equipment and contact technical
service.
COOLING SYSTEM
LOW IN GAS
Tap on icon
in order to
view alarm list and to mute it
(par.5.2.16)
Turn the refrigerator off from
the display, -STAND BY-
Par.5.2.8. Turn it back on
after a few seconds.
Should this alarm or alarms
related to temperature occur
again, contact technical
service.
Leakage or micro-leakage in the
cooling system
POWER SUPPLY
Turn the refrigerator off from
the display, -STAND BY-
Par.5.2.8. After a while turn it
on and the visual and
acoustic alarm is off.
Tap the off alarm/fault icon so
to permanently delete the
signal (par.5.2.16).
If the problem occurs again,
contact technical service
Electrical parts or electronic board
fuses failure (n.1 5x20 2A o n.1
5x20 4A).
NO BATTERY
or
BATTERY FAILURE
Batteries disconnected
Weak batteries
Check the correct connection of
the batteries.
Replace batteries (par.8.5)
Re-insert SD card
Purchase new SD card (par.)
Insert new SD card
SD absent
Damaged SD
You have inserted a SD of
another refrigerator
MAINS FAILURE
Do check the electrical
connections by a qualified
electrician
If the problem occurs again,
contact technical service
Temporary black out >10 mins