Instructions, Maintenance and Service Manual ULTRAFREEZER Series
12
7.10. Resetting alarm/buzzer
If there is an alarm, a message appears on the display and a LED light on (with the exception of
the alarm temperature relay). At the same time, the potential free contact and the buzzer is
activated. The buzzer can be reset by pressing the
‘MUTE’
button. When the alarm situation is
solved by the equipment, the buzzer, the relay and the display will be reset automatically.
Despite this the original alarm remains in the display, to show that there has been an alarm.
Reset alarm
Press
Code
.
Press
▲
= 1;
Setp.
= 2;
Batt.
= 3,
▼
= 4, the password.
Press
Setp.
to reset alarm.
If a temperature sensor is defective during installation, the controller does not work properly
and the sensor needs to be replaced. If the sensor is defective after some time of operation, the
controller shall switch the cooling on and off, depending on the previous ON/OFF switching
times of the cooling cycles. In this way, the temperature in the freezer remains. Of course this
situation needs to be repaired as soon as possible.
7.11. Storage of the log on a USB stick
To download the log files on the USB pen, put the USB pen (with sufficient free space) into the
USB connection at the front of the freezer (behind the condenser plate, which can be removed
very easily).
Enter the password and press on the
▲
button after having entered the password. When the
text
L.USb
appears on the screen, the ‘
Setp.’
button needs to be pressed to start up the storing.
The percentage of loading is shown in the screen. After the transfer of all the log files, the text
“
donE
” will appear on the display. There are two error codes possible:
E.US1
, no USB stick has
been inserted.
E.US2
, there is not enough free space on the stick.
To check the name of the file (MAC-address), please do the following:
If the correct password is used, press the
‘PRG’
button.
Press
▲
button, the display will show
P91
.
By pressing
‘SETP.’
the display will show a
‘MAC address’
.
Write this number down.
The display returns to its normal status when the
‘PRG’
button is pressed or when
during 30 seconds no button is used.
7.12. Open log
The log (with the correct MAC-address) on the USB stick can be opened by using the pc program
sl_log.exe (which can be downloaded). After opening the program, the following screen
appears.