6-2
Isotemp Basic
Fisher Scientific
Section 6
Specifications
IE1786V (8958)
-50C to -86C (-58F to -123F) in an 18C to 32C* (64.4F to 89.6F) ambient
33.3”W x 77.9”H x 38,9”
84.6 x 197.9 x 94.0cm
23.0”W x 51.5”H x 25.3”
58.4 x 130.8 x 64.3cm
17.3 cu. ft. (489.9 liters)
Two 1 HP (2545 BTUH each)
Non-CFC, foamed-in-place urethane: 5.0” (12.7 cm) cabinet; 4.5” (11.4 cm) door
230V, 50 Hz, 12.0FLA
Operating Range: 208-240V
15 amp Dedicated Circuit,
15 Amp Time Delay Breaker
830 lbs. (376 kg)
IE1786D (8957)
33.3”W x 77.9”H x 38.9”
84.6 x 197.9 x 94.0cm
23.0”W x 51.5”H x 25.3”
58.4 x 130.8 x 64.3cm
17.3 cu. ft. (489.9 liters)
208-230V, 60 Hz, 12.0FLA
Operating Range: 208-240V
15 amp Dedicated Circuit,
15 Amp Time Delay Breaker
830 lbs. (376 kg)
IE1786A (8956)
33.3”W x 77.9”H x 38.9”
84.6 x 197.9 x 94.0cm
23.0”W x 51.5”H x 25.3”
58.4 x 130.8 x 64.3cm
17.3 cu. ft. (489.9 liters)
120V, 60 Hz, 16.0 FLA
Operating Range: 108-130V
20 amp Dedicated Circuit,
20 amp Time Delay Breaker
830 lbs. (376 kg)
Catalog No. (Model)
Temperature Range
Exterior Dimensions
Interior Dimensions
Capacity
Refrigeration
Insulation
Electrical -
nominal voltage ±10%
Breaker
Requirements
Shipping Weight:
Motor
*
Compressors may not cycle off with cabinet running at -86C in a 32C ambient.
IE2386 A (8964)
-50C to -86C (-58F to -123F) in an 18C to 32C* (64.4F to 89.6F) ambient
40.8”W x 77.9”H x 38.9”
103.6 x 197.9 x 94.0cm
30.6”W x 51.5”H x 25.3”
77.7 x 130.8 x 64.3cm
23.0 cu. ft. (651.3 liters)
Two 1 HP (2545 BTUH each)
Non-CFC, foamed-in-place urethane: 5.0” (12.7 cm) cabinet; 4.5” (11.4 cm) door
120V, 60 Hz, 16.0 FLA
Operating Range: 108-130V
20 amp Dedicated Circuit,
20 amp Time Delay Breaker
900 lbs. (408 kg)
IE2386D (8963)
40.8”W x 77.9”H x 38.9”
103.6 x 197.9 x 94.0cm
30.6”W x 51.5”H x 25.3”
77.7 x 130.8 x 64.3cm
23.0 cu. ft. (651.3 liters)
208-230V, 60 Hz, 12.0FLA
Operating Range: 208-240V
15 amp Dedicated Circuit,
15 Amp Time Delay Breaker
900 lbs. (408 kg)
IE2386V (8959)
40.8”W x 77.9”H x 38.9”
103.6 x 197.9 x 94.0cm
30.6”W x 51.5”H x 25.3”
77.7 x 130.8 x 64.3cm
23.0 cu. ft. (651.3 liters)
230V, 50 Hz, 12.0FLA
Operating Range: 208-240V
15 amp Dedicated Circuit,
15 Amp Time Delay Breaker
900 lbs. (408 kg)
Catalog No. (Model)
Temperature Range
Exterior Dimensions
Interior Dimensions
Capacity
Refrigeration
Insulation
Electrical-
nominal voltage ±10%
Breaker
Requirements
Shipping Weight:
Motor