3-2
|
Isotemp
Fisher Scientific
Chapter 3
|
Installation
Electrical Requirements
Refer to the bath's nameplate for specific electrical requirements.
The bath is intended for use on a dedicated outlet.
The bath construction provides protection against the risk of electrical shock by
grounding appropriate metal parts. The protection will not function unless the power
cord is connected to a properly grounded outlet. It is the user's responsibility to
assure a proper ground connection is provided.
The circuit protector on the rear of the bath is designed to protect the bath's internal components.
Note
If the circuit protector activates allow the temperature to cool before resetting. Restart the
bath. Contact us if the circuit protector activates again.
The bath's electrical power cord is the disconnecting device, it must be easily
accessible at all times.
Ensure the cord does not come in contact with the reservoir contents.
Note
Before inserting the electrical cord into the bath connection ensure the circuit protector is in
the
O
(off) position.
Once the cord is connected to the bath, connect the other end to the main power source.
DANGER
CAUTION
CAUTION