3.2.2.2 Degassing the System
1) Press the UP button to start a 60-second cleaning cycle countdown. The
UP indicator light blinks if UP water quality does not meet the preset
standard.
2) After the cleaning cycle, the system begins dispensing water.
During the cleaning, RO indicator light blinks; indicating water quality does
not meet the preset standard. Once the cleaning is completed, RO
indicator light stops blinking and water quality reaches the standard.
3) Discharge the initial 5 - 10L of water (about 10 - 15 minutes) after installing
a new purification cartridge. Press the UP button again to stop dispensing.
14/10/17 14:59
RO: 12.0
µ
S/cm 25
UP: 1.0 M
Ω
.cm 25
LOW PRESSURE