 18L low temp incubator         3-1

3

Installation

Environmental conditions

The Fisherbrand 18L low temp incubator device must only be operated in a location that 
meets all of the specific ambient condition requirements listed below:

Location requirements

• Dry place of erection within a closed space which is free from droughts.

• Dust loading should be in keeping with the degree of contamination, which is Degree 2 

according to IEC EN 61010-1. Using the incubator in an atmosphere with electrically 
conductive dust is prohibited.

• The minimal distance to adjacent surfaces must be observed on all sides (see Section 3.3).

• The operating room must be equipped with appropriate ventilation.

• Solid, level, fire-proof surface.

• Vibration-proof substructure (floor stand, lab table) capable of bearing the dead weight of 

the device and its accessories (particularly if two devices are stacked).

• The electrical circuitry of the device has been designed for an operating height of up to 

2000 m above sea level. 

• The permissible ambient temperature is b18 °C and 32 °C (+64.4 °F and + 

89.6 °F) for an incubation temperature of between 17 °C and 40 °C (62.6 °F and 104 °F).

• Relative humidity up to 80% (maximum; preferably 60-70%), non condensing. 

• Dew is to be avoided, e.g. after changes of location or transportation. Should 

condensation exist, wait until the moisture has evaporated completely before connecting 
the device to a power source and powering up.

• Avoid direct exposure to sunlight.

• Devices that produce excessive amounts of heat must not be placed near the device.

• Storage temperature: When the device is placed in intermediate storage, make sure that 

the ambient temperature is between 20 °C and 60 °C (68 °F and 140 °F) and the 
maximum relative humidity does not exceed 90%, non-condensing. 

Summary of Contents for 50152744

Page 1: ......

Page 2: ...c 300 Industry Drive Pittsburgh Pennsylvania 15275 United States Thermo Fisher Scientific provides this document to its customers with a product purchase to use in the product operation This document...

Page 3: ...ntal conditions3 1 Room ventilation3 2 Installation3 2 Transport3 2 Chapter 4 Description of unit4 1 Device overview4 1 Mains connection4 2 Useable space components4 2 Condensation tray4 3 Chapter 5 I...

Page 4: ...and disinfection9 1 Empty the condensation tray 9 1 Cleaning exterior surfaces9 1 Wipe spray disinfection9 1 Chapter 10 Maintenance10 1 Inspections and checks10 1 Chapter 11 Disposal11 1 Overview of...

Page 5: ...es targeted at personnel working with the subject matter device detailing the disinfection measures to be employed for the device and the accessories used with it not to be used in a damp atmosphere t...

Page 6: ...condition that the device is operated and serviced exclusively in accordance with its intended purpose and as described in these operating instructions the device is not modified only original spare...

Page 7: ...plication Wear safety gloves Wear safety goggles Harmful liquids Electrical shock Fire hazard Contamination hazard Observe operating instructions Mark of conformity USA Canada This device has been tes...

Page 8: ...s which these cultures require to be simulated due to it exactly controlling the useable space temperature The device is only to be used in closed spaces Incorrect use To avoid the risk of explosion d...

Page 9: ...ng film Polyethylene Acceptance inspection After the device has been delivered check the delivery immediately for completeness possible damage If components are missing or damage is found on the devic...

Page 10: ...Delivery of the device Scope of supply 2 2 18L low temp incubator Fisherbrand...

Page 11: ...f bearing the dead weight of the device and its accessories particularly if two devices are stacked The electrical circuitry of the device has been designed for an operating height of up to 2000 m abo...

Page 12: ...tional ventilation may have to be provided as necessary The room ventilation is to be of a technical form which is designed in accordance with the thermal input Installation When installing the device...

Page 13: ...low temp incubator 4 1 4 Description of unit Device overview Front view 1 2 4 3 Abbildung 4 1 Front view Fisherbrand 18L low temp incubator 1 Temperature display 2 Control panel 3 Door door handle 4 P...

Page 14: ...duced to a minimum and therefore supports the protection against contamination and its easy and efficient elimination All of the parts of the inner container of the Fisherbrand 18L low temp incubator...

Page 15: ...3 Condensation tray A removable tray Abb 4 3 2 is provided underneath the device which collects the condensation which collects whilst the useable space is in use Abbildung 4 3 Cooling or heating ele...

Page 16: ...Description of unit Condensation tray 4 4 18L low temp incubator Fisherbrand...

Page 17: ...lates Insert the perforated plates in the grooved profiles on the side walls of the useable space Preparing the useable space The Fisherbrand 18L low temp incubator is not delivered in a sterile condi...

Page 18: ...e check to see if the power supply voltage corresponds with the specifications on the nameplate on the rear of the device If the voltage V and current A ratings given are not as required do not connec...

Page 19: ...rectly inserted Starting Operation 1 Switch the device on with button 2 The display shows the current useable space temperature 3 Set the target values for the temperature using buttons and on the con...

Page 20: ...Operation Starting Operation 6 2 18L low temp incubator Fisherbrand...

Page 21: ...or Tabelle 7 1Operating elements of the Fisherbrands 18L low temp incubator No Designation Function 1 Display with permanent indication of the current useable space temperature in C in two figures a s...

Page 22: ...perature You can preselect temperatures between 15 C and 40 C 59 F and 104 F The thermal output and cooling capacity can be influenced by the ambient temperature the number of samples which are to be...

Page 23: ...he environment of the incubator 1 Remove all items and all auxiliary agents from the useable space 2 Turn the device off using the control panel 3 Unplug the power connector and secure it against acci...

Page 24: ...Shut down Shutting the device down 8 2 18L low temp incubator Fisherbrand...

Page 25: ...g and disinfection work Switch the device off with button Remove the plug from the socket and secure it against reconnection Check whether the device is voltage free Cleaning exterior surfaces Remove...

Page 26: ...behind the door seal Wipe the display dry with a rag made of 100 microfiber Alcoholic disinfectants Disinfectants having an alcohol content of more than 10 may form in combination with air easily com...

Page 27: ...samples from the useable space and store them in a safe place 2 Spray or wipe over the surfaces of the useable space and or the perforated plates with a disinfectant 3 Allow time for the disinfectant...

Page 28: ...or Fisherbrand Final disinfection 1 Again spray or wipe over the surface of the useable space and the removed perforated plates with a disinfectant 2 Allow time for the disinfectant to act as specifie...

Page 29: ...d below must be checked at regular intervals Daily check Check the water level in the condensation tray underneath the device on a daily basis In order to empty it pull the condensation tray completel...

Page 30: ...Maintenance Inspections and checks 10 2 18L low temp incubator Fisherbrand...

Page 31: ...mination of the device and its components Prior to disposal it is therefore mandatory that all incubator com ponents be properly decontaminated Clean the incubator components thoroughly then disinfect...

Page 32: ...with diverse plastics on epoxy resin bonded printed circuit boards Plastic components general see material labeling Exterior housing Plastic Device rear panel Plastic Outer door Plastic Heater Peltie...

Page 33: ...mption 0 45 Amps Protection Installation class II Degree of protection IP20 Max mechanical load 4 41 lb per plate 11 02 lb per device Connection voltage 100 240V AC 10 50 60 Hz Connection with the enc...

Page 34: ...8 F at 98 6 F 0 2 C at 37 C 32 36 F at 98 6 F Preheat times device empty to 98 of the working tem perature 37 C 98 6 F 15 minutes Recovery times device empty door 30 s open to 98 of the working temper...

Page 35: ...Fisherbrand 18L low temp incubator 13 1 13 Accessories Material No Description 10031 744 Stainless steel perforated shelf for small cooled incubator 18L...