Installation
Room ventilation
3-2 18L low temp incubator
Fisherbrand
Room ventilation
Heat dissipating from the device during continuous operation may cause a change in the
room climate.
Therefore, the device must only be installed in rooms with sufficient ventilation.
Do not install the incubator in room recesses without ventilation.
When several devices are to be placed in the same room, additional ventilation may have to
be provided as necessary.
The room ventilation is to be of a technical form which is designed in accordance with the
thermal input.
Installation
When installing the device, make sure that the installation and supply connections remain
freely accessible. The minimum wall clearance is 5 cm / 2 in on the sides and back. No other
devices are to be placed on the device.
The cooling fan on the back of the device is not to be covered.
It is not to suck-in any loose items (e.g. paper).
Transport
The device is not to be lifted by the door or attachments when being transported.