Cleaning and disinfection
Wipe / spray disinfection
9-2 18L low temp incubator
Fisherbrand
Space is to be made available for this disinfection work which is suitable for a treatment of
the removed device components with cleaning agents and disinfectants.
Incompatible cleaners
Some device components are made of plastics. Solvents may at-
tack plastics. Strong acids or alkaline solutions may cause embrit-
tlement of plastics.
Moisture-sensitive components
Do not spray cleaning agent onto the display.
When wiping over, ensure that no moisture is able to penetrate the
display or run behind the door seal.
Wipe the display dry with a rag made of 100% microfiber.
Alcoholic disinfectants!
Disinfectants having an alcohol content of more than 10% may
form, in combination with air, easily combustible and explosive gas
mixtures.
When using such disinfectants, avoid open flames or exposure to
excessive heat during the entire disinfection process!
• Use such disinfectants only in adequately ventilated rooms.
• After the disinfectant has been allowed to react, wipe the
cleaned incubator components thoroughly dry.
• Observe safety regulations to avoid fire and/or explosion
hazard caused by alcohol-containing disinfectants.
Chloride-containing disinfectants!
Chloride-containing disinfectants can corrode stainless steel.
• Use only disinfectants that do not affect stainless steel!
Summary of Contents for 50152744
Page 1: ......
Page 10: ...Delivery of the device Scope of supply 2 2 18L low temp incubator Fisherbrand...
Page 16: ...Description of unit Condensation tray 4 4 18L low temp incubator Fisherbrand...
Page 20: ...Operation Starting Operation 6 2 18L low temp incubator Fisherbrand...
Page 24: ...Shut down Shutting the device down 8 2 18L low temp incubator Fisherbrand...
Page 30: ...Maintenance Inspections and checks 10 2 18L low temp incubator Fisherbrand...