Health hazard
The surfaces of the useable space may be contaminated. Contact
with contaminated cleaning liquids may cause infections. Disin-
fectants may contain harmful substances.
When cleaning and disinfecting, always observe the safety in-
structions and hygiene guidelines!
• Wear safety gloves.
• Wear safety goggles.
• Wear mouth and respiratory system protection gear to protect
your mucous membranes.
• Observe the safety instructions of the disinfectants
manufacturer and the hygiene supervisor.
Cleaning and disinfection
Wipe / spray disinfection
Fisherbrand
18L low temp incubator 9-3
Predisinfection
1. Remove all samples from the useable space and store them in a safe place.
2. Spray or wipe over the surfaces of the useable space and or the perforated plates with a
disinfectant.
3. Allow time for the disinfectant to act as specified by the manufacturer.
Moisture-sensitive components
Do not spray disinfectants onto the rear of the interior.
Cleaning
1. Remove the perforated plates from the useable space.
2. Wipe over the surfaces of the useable space and the removed perforated plates with tepid
water containing household washing-up liquid. Also remove all stubborn stains with warm
water and washing-up liquid without residues.
3. Rinse the cleaned surfaces 3-5 times with autoclaved water so as to fully remove cleaning
agent residues.
4. Then rub the surfaces and cleaned perforated plates dry with a soft sterile cloth.
Summary of Contents for 50152744
Page 1: ......
Page 10: ...Delivery of the device Scope of supply 2 2 18L low temp incubator Fisherbrand...
Page 16: ...Description of unit Condensation tray 4 4 18L low temp incubator Fisherbrand...
Page 20: ...Operation Starting Operation 6 2 18L low temp incubator Fisherbrand...
Page 24: ...Shut down Shutting the device down 8 2 18L low temp incubator Fisherbrand...
Page 30: ...Maintenance Inspections and checks 10 2 18L low temp incubator Fisherbrand...