Technical Data
12-2 18L low temp incubator
Fisherbrand
* The national electrical engineering rules and corresponding technical connection
conditions are to be adhered to when connecting to the mains supply.
Temperature deviations
- spatial with an ambient tem-
perature of 23 °C (73.4 °F)
- temporal with an ambient
temperature of 23 °C
(73.4 °F)
+/- 1 °C at 37 °C (+/- 33.8 °F at 98.6 °F)
+/- 0.2 °C at 37 °C (+/- 32.36 °F at 98.6 °F)
Preheat times (device empty,
to 98% of the working tem-
perature)
37 °C (98.6 °F) - 15 minutes
Recovery times (device
empty, door 30 s open, to
98 % of the working tempera-
ture)
37 °C (98.6 °F) - 5 minutes
On site single fuses*
- safety fuse (time-lag) T 16 A or circuit breaker B
16
- connection via a fault current circuit breaker
(triggering current < 30 mA) is recommended
Controllable temperature
range in the useable space
+17...40 °C (+62.6...104 °F) with an ambient tem-
perature of between 18 and 32 °C (64.4 and
89.6 °F)
Designation
Value