5 BASIC TROUBLESHOOTING
Problem
Possible Cause
Solution
System inactive
No power supply
Main fuse blown
Circuit partially broken
Check the power supply
Check the main fuse
Contact a ULS agent
No water
Gas in the system
Pump failure
Degassing the system
Contact a ULS agent
Low water flow rate
Final filter clogged
System internal water
flow clogged
Replace final filter
Contact a ULS agent
If problem occurring is not listed in the chart above, please contact Fisher