13
QC Lockout
Select
QC Lockout
and press
ENTER
to cycle the options
Off
or
On
.
Off
disables the QC lockout function. The
QC
symbol and Setup Options are not
displayed.
On
enables the QC lockout function. The
QC
symbol is displayed. QC Lockout
timing can be set for
Every 8 Hours,
Daily, Weekly
or
Monthly
.
Time
Select
Time
and press
ENTER
to set the QC test for
Every 8 Hours,
Daily,
Weekly
or
Monthly
.
Time Setup
Select
Time Setup
and press
ENTER
.
Every 8 Hours / Daily Weekly / Monthly
The first highlighted digit will be changed. Press
▲
or
▼
to change. Continue to
press
▲
or
▼
until the correct number is shown. Press
◄
or
►
to progress to the
next digit to be changed. Press
▲
or
▼
to increase or decrease the number.
Change each of the time and date digits/fields in this manner until the desired date
is shown. Press
ENTER
to save the changes and exit the screen. Press
CANCEL
to not accept changes and exit.
There is no
Date Setup
if QC tests are set for
Every 8 Hours
or
Daily
.
For the
Monthly
option,
Date
can be set from 01 to 28. 29, 30
and 31 are invalid.
Run QC Test
Select
Run QC Test
and press
ENTER
to perform a QC test before the next
scheduled test time.
When a QC test is run for the first time, the analyzer will show the screen below if
QC values have not been set up.