26
Press
START
to begin strip testing. The testing screen will be displayed as
shown below. This screen example shows a total of 10 strips to be run. It asks
for the first strip to be placed on the Strip Holder.
The next screen shows the barcode of the sample to be run and the next sample
ID to be run. This reduces the possibility of error and mismatch between the
barcode and the sample to be run.
When the test is complete, the
Results Screen
will show briefly. The next strip
can be placed on the Strip Holder.
Stop testing by pressing
CANCEL
when the prompt appears. All other testing
procedures are the same as testing with no barcode.
Urine Controls QC Testing
Ensure the operating
Mode
is set to
QC
. All test numbers in QC mode will begin
with
2
. This allows results to be searched for and found easily.
Strip Preparation
Allow the strip and urine controls to reach room temperature at 15-30ºC (59-86ºF)
prior to testing.
Note:
Use
AimStrip
®
and
Fisherbrand
®
Urinalysis Reagent Strips for
proper functioning and accurate results.