27
Remove strips from the closed canister and use them as soon as possible. Close
the canister tightly immediately after removing the strips.
Urine Control Test Procedures
The urine control test procedures are the same as Normal Operation, no Barcode
reader, or Normal Operation, Barcode reader installed. Refer to Section 5
Analyzer Operation.
The results obtained during the Quality Control test will be referenced by the
analyzer with pre -programmed target values and report generated, If
Auto-print
is
set to
On
, the result will be printed. If the QC test "Fails", please contact
Germaine Laboratories, Inc. Technical Support.
QC lockout
When
QC lockout
is turned
On
, the user will be notified when a QC test is required.
The following screen will be displayed. The
QC
symbol will flash.
Note:
Use
AimStrip
®
and
Fisherbrand
®
Urinalysis Reagent Strips for
proper functioning and accurate results.
If it is the first time to run the QC test, refer to the section on
Setting the QC Values
to set up QC test values.
Press
CANCEL
to return to the Initial Screen. If canceled, the analyzer will
change the test mode to
STAT
. An “E” will be displayed after the test number on
all printouts to show QC tests were out of date and pending.
Press
ENTER
to run QC tests.