Figure 3. HT Cell Seeders

 in a 24-well HT Baseplate. 

 

I

NSTRUCTIONS

 

 

1.

 

Insert the HT Cell Seeder

 into a 24-well HT Baseplate, similar to how the cylindrical loading 

stations are placed in the baseplate.  

2.

 

Place an HT BioFlex

® 

plate and its respective gasket over the HT Cell Seeder

, ensuring that the 

wells in the plate align with the posts on the HT Cell Seeder

3.

 

Using the FX-5000

 Tension System, create and start a regimen with the following settings: 

 

SHAPE: Static 

 

MIN: 0.0 

 

MAX: 8.0 

 

FREQ: 1 Hz (this value must still be entered even though the regimen is static) 

 

DURATION: Equal to the seeding time (see step 5 below) plus the time needed to plate 
the cells into the well. 

 

PLATFORM: HT 24-Well Plate (Cylindrical LS)  

 

NOTE: The membrane may be subjected to strains up to 1-2% (10 to 20 microns per 1 millimeter) when 
seeding, due to the inherent strain caused when using the HT Cell Seeder

.  

 

4.

 

The recommended suspension volume when using an HT Cell Seeder

 for an individual well is 

175 µL. This volume is large enough to allow uniform distribution of the media-cell suspension 
within the well when dispensing with a 1000 µL micropipette (

Figure 4

). Also, this volume is 

small enough to prevent the media-cell suspension from spilling out of the HT Cell Seeder

 

when handling the baseplate. 