18 Outputs
122
UMFLUXUS_F8V4-6-1EN, 2018-10-10
18.1.1
Output Range
18.1.2
Error Value
In the following dialog, an error value can be defined which is output in case a source item cannot be measured, e.g., if
there are gas bubbles in the fluid.
Miscellaneous
c-Medium
sound speed of the fluid
Concentration
concentration of the fluid
Signal
signal amplitude of a measuring channel
SCNR
ratio useful signal to correlated disturbance signal
VariAmp
standard deviation of the signal amplitude
Density
density of the fluid
Pressure
pressure of the fluid
When configuring an analog output, the output range will be defined now. Select a list item
or
other range
... to enter the output range manually.
If
other range...
is selected, enter the values
Output MIN
and
Output MAX
. Press
ENTER after each input.
This error message will be displayed if the output range is not min. 10 % of the max. output
range. The next possible value will be displayed. Repeat the input.
example: I
MAX
- I
MIN
≥
2 mA for a 4…20 mA current output
Tab. 18.2: Error output
error value
results
Minimum
the lower limit of the output range is transmitted
Hold last value
the last measured value is transmitted
Maximum
the upper limit of the output range is transmitted
Other value...
The value has to be entered manually. It has to be within the limits of the output.
Tab. 18.1: Configuration of the outputs
source item
list item
output
I1 Output range
↕
4/20 mA
I1 Output MIN
↕
10.0 mA
I1 Output MAX
↕
11.0
mA
I1 Output MAX
↕
12.0 MINIMAL