A Menu Structure
FLUXUS ADM 8027, F801, ADM 8127B
UMFLUXUS_F8V4-6-1EN, 2018-10-10
137
input of the number of sound paths
display of the transducer distance to be adjusted between the inner edges of
the transducers
Program Branch
Output Options
main menu: selection of the program branch
Output Options
selection of the channel whose output options are to be defined
selection of the physical quantity
selection of the unit of measurement for the physical quantity
input of the duration over which a floating average of the measured values
has to be determined
range: 1...600 s
activation of the data logger
activation of the measured values transmission to a PC or a printer via a seri-
al interface
selection of the storage rate for storing measured values in the data logger
This display will only be indicated if
Output Options\Store Meas.Data
and/or
Serial Output
are activated.
input of the storage rate if
Storage Rate
=
EXTRA
is selected
range: 1...43 200 s (= 12 h)
Current loop
activation of a current output
This display will only be indicated if the current output has been installed in
Special Funct.\SYSTEM settings\Proc. outputs
.
selection whether the sign of the measured values is to be considered for the
output
This display will only be indicated if
Current Loop
is activated.
input of the lowest/highest measured value to be expected for the current out-
put
The values are assigned to the lower/upper limit of the output range.
These displays will only be indicated if
Current Loop
is activated.
INIT-resis-
tant
A: Sound Path
2
NUM
Transd. Distance
A:54 mm Reflex
par mea >OPT< sf
Output Options
Output Options
↕
for Channel
A:
Physic. Quant.
↕
Volume flow
Volume in:
↕
m3/h
Damping
10
s
Store Meas.Data
no
>YES<
Serial Output
no
>YES<
Storage Rate
↕
Once per 10 sec.
Storage Rate
1
s
Current Loop
I1: no
>YES<
Meas.Values
>ABSOLUT<
sign
Zero-Scale Val.
0.00
m3/h
Full-Scale Val.
300.00
m3/h