FLUXUS ADM 8027, F801, ADM 8127B
158
UMFLUXUS_F8V4-6-1EN, 2018-10-10
D
Maintenance
The transmitter and the transducers are practically maintenance-free.
D.1
Cleaning
• Clean the transmitter with a soft cloth. Do not use detergents.
• Remove traces of acoustic coupling compound from the transducers with a paper tissue.
D.2
Examination of the O-ring
• Check the O-ring when opening the flameproof enclosure housing (FLUXUS ADM 8027, see Fig. D.1, FLUXUS F801,
Flameproof enclosure
• If the O-ring is defective or the housing does not seal anymore, change it (see Tab. D.1).
Housing with increased safety
• If the O-ring is defective or the housing does not seal anymore, have it replaced.
Attention!
O-rings on increased safety housings may only be replaced by trained FLEXIM personnel.
Note!
Store the O-ring replacement in a hermetically sealed and dark place.
Fig. D.1: FLUXUS ADM 8027
Fig. D.2: FLUXUS F801
flameproof enclosure
housing
with increased
safety
housing with increased
safety
O-ring
O-ring
flameproof enclosure
housing
with increased
safety
O-ring